Zpravodaj Hnědé uhlí 2024, 64(1)
Uncertainty estimation and compliance test with the reference method of Fidas 200 E analysers in the measurement of suspended particles PM10
FIDAS 200 E optoelectronic analysers are newly used in the measurement network of VúHU a.s. Before their inclusion in the measurement network, a compliance test was performed with the reference gravimetric method according to ČSN EN 12341. When evaluating the results, the statistical method specified in ČSN EN 16450 was applied. The determination results of the uncertainties were in accordance with the requirements of Decree No. 330/2012 Coll., on the method of evaluating the level of pollution, the extent of informing the public about the level of pollution and, in smog situations, the uncertainty for measuring PM10 particles.
Keywords: suspended particles PM10, FIDAS 200, E - Leckel
Published: March 1, 2024 Show citation